Spectral CT imaging as a new quantitative tool? Assessment of perfusion defects of pulmonary parenchyma in patients with lung cancer
Abstract
Objective: This study investigated the capability of dual-energy spectral computed tomography (CT) to quantitatively evaluate lung perfusion defects that are induced by central lung cancer.
Methods: Thirty-two patients with central lung cancer underwent CT angiography using spectral imaging. A univariate general linear model was conducted to analyze the variance of iodine concentration/CT value with three factors of lung fields. A paired t-test was used to compare iodine concentrations and CT values between the distal end of lung cancer and the corresponding area in the contralateral normal lung.
Results: Iodine concentrations increased progressively in the far, intermediate and near ground sides in the normal lung fields at 0.60±0.28, 0.93±0.27 and 1.25±0.38 mg/mL, respectively (P<0.001). The same trend was observed for the CT values [–(840.64±49.08), –(812.66±50.85) and –(760.83±89.17) HU, P<0.001]. The iodine concentration (0.70±0.42 mg/mL) of the lung field in the distal end of lung cancer was significantly lower than the corresponding area in the contralateral normal lung (1.19±0.62 mg/mL) (t=–7.23, P<0.001). However, the CT value of lung field in the distal end of lung cancer was significantly higher than the corresponding area in the contralateral normal lung [–(765.29±93.34) HU vs. –(800.07±76.18) HU, t=3.564, P=0.001].
Conclusions: Spectral CT imaging based on the spectral differentiation of iodine is feasible and can quantitatively evaluate pulmonary perfusion and identify perfusion defects that are induced by central lung cancer. Spectral CT seems to be a promising technique for the simultaneous evaluation of both morphological and functional lung information.
Methods: Thirty-two patients with central lung cancer underwent CT angiography using spectral imaging. A univariate general linear model was conducted to analyze the variance of iodine concentration/CT value with three factors of lung fields. A paired t-test was used to compare iodine concentrations and CT values between the distal end of lung cancer and the corresponding area in the contralateral normal lung.
Results: Iodine concentrations increased progressively in the far, intermediate and near ground sides in the normal lung fields at 0.60±0.28, 0.93±0.27 and 1.25±0.38 mg/mL, respectively (P<0.001). The same trend was observed for the CT values [–(840.64±49.08), –(812.66±50.85) and –(760.83±89.17) HU, P<0.001]. The iodine concentration (0.70±0.42 mg/mL) of the lung field in the distal end of lung cancer was significantly lower than the corresponding area in the contralateral normal lung (1.19±0.62 mg/mL) (t=–7.23, P<0.001). However, the CT value of lung field in the distal end of lung cancer was significantly higher than the corresponding area in the contralateral normal lung [–(765.29±93.34) HU vs. –(800.07±76.18) HU, t=3.564, P=0.001].
Conclusions: Spectral CT imaging based on the spectral differentiation of iodine is feasible and can quantitatively evaluate pulmonary perfusion and identify perfusion defects that are induced by central lung cancer. Spectral CT seems to be a promising technique for the simultaneous evaluation of both morphological and functional lung information.